• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MMJ SA DE CV (USD) SHILEY; TUBE TRACHEOSTOMY AND TUBE CUFF Back to Search Results
Model Number 6CFS
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 11/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the device had extreme difficulty in locking and unlocking the inner cannula of tracheostomy tube.There was no patient involvement.
 
Additional Manufacturer Narrative
Evaluation summary: one sample of device was received for evaluation.The received components are one cuffless tracheostomy outer cannula, one inner can nula with integral 15-mm twist-lock connector, one inner cannula with closed lumen and low-profile connector, one cannula with open lumen and low-profile connector and one obturator, all the received components are size 6.A functional test was performed inner cannulas were inserted and locked into the outer cannulas and no difficulty was observed at the time of locking and unlocking the cannula.A torque test was performed using the inner cannula with integral 15-mm twist-lock connector, it was measured 1.68 lbs/in (spec 0.63 to 2.5 lbs/in.Information has been added to the database and trends will continue to be monitored medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE TRACHEOSTOMY AND TUBE CUFF
Manufacturer (Section D)
MMJ SA DE CV (USD)
ave henequen no 1181 desarroll
ciudad juarez,mx 32590
MX  32590
MDR Report Key9521881
Report Number2936999-2019-01060
Device Sequence Number10885538
Product Code JOH
UDI-Device Identifier60884522006960
UDI-Public60884522006960
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K962173
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/19/2024
Device Model Number6CFS
Device Catalogue Number6CFS
Device Lot Number19E0457JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2020
Initial Date Received by Manufacturer 12/11/2019
Supplement Date Received by Manufacturer01/20/2020
Initial Report FDA Received Date12/26/2019
Supplement Report FDA Received Date02/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-