| Model Number |
24716 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/18/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in mid to distal anterior tibial artery.A 2.5x220x150 (4f) sterling balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured before reaching the rated burst pressure.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.There was blood in the inflation lumen and balloon.Microscopic examination revealed a pinhole over the proximal markerband.Microscopic examination of the balloon presented no irregularities in the balloon material or the ro marker that could have contributed to the damage.There was no other damage revealed.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon rupture occurred.The target lesion was located in mid to distal anterior tibial artery.A 2.5x220x150 (4f) sterling balloon catheter was advanced for dilatation.However, during inflation, the balloon ruptured before reaching the rated burst pressure.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was stable.
|
| |
|