Model Number PCB00
Medical Device Problem Codes Unintended Movement (3026); Separation Problem (4043)
Health Effect - Clinical Codes Corneal Edema (1791); No Code Available (3191)
Date of Event 11/07/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/date of birth: unknown, information not provided.(b)(6).(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) had haptic missing.Additional information was received stating that when the pcb00 lens was implanted into the eye, they noticed that one haptic may be missing.After three to four weeks, patient came back as the lens was not staying in position properly, the lens was explanted/exchanged.The patient was worried, but surgery was uncomplicated.There was no suture or vitrectomy, however, an incision enlargement was required.Also noted is neuro-128 onguent was prescribed by doctor due to corneal edema persistent one-month post-op.Visual acuity pre-op (pre-initial implant): 20/200.Visual acuity post-op (post-initial implant): 20/200.Visual acuity post-op (post-replacement): 20/70 (one month post first surgery).No further information provided.
 
Additional Manufacturer Narrative
Additional information: device available for evaluation; returned to manufacturer on 2/10/2020.Device evaluation: the product was received at site inside a bag containing a plastic identified container with the lens inside.The lens was evaluated and the following were the observations: lens cut and haptic detached.The other haptic looks in good conditions.The complaint issue was verified, however due the condition of the returned (handled) a product deficiency and malfunction could not be determine.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.A search revealed no additional investigation requests received for this production order number.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9522337
Report Number2648035-2019-01382
Device Sequence Number16869639
Product Code HQL
UDI-Device Identifier05050474558397
UDI-Public(01)05050474558397(17)211108
Combination Product (Y/N)Y
Initial Reporter CountryCA
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 11/05/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/08/2021
Device Model NumberPCB00
Device Catalogue NumberPCB0000265
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 11/28/2019
Supplement Date Received by Manufacturer01/29/2020
10/25/2020
Initial Report FDA Received Date12/26/2019
Supplement Report FDA Received Date02/24/2020
11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-