Model Number 16-02-80
Device Problem Contamination (1120)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/22/2019
Event Type  Malfunction  
Manufacturer Narrative

There was no known patient involvement. The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k052601). Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices. The z number is z-2076/2081-2015. Livanova (b)(4) manufactures the heater-cooler system 3t. The incident occurred in (b)(6). A review of the dhr could not identify any deviations or nonconformities relevant to the issue. Through follow up communication, livanova (b)(4) learned that the device is positioned inside the operation room, away from the patient with an estimated distance of approximately 4 meters to the patient. A lab report stating contamination with mycobacterium chimaera was provided. Corrective actions are in progress for this issue. If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.

 
Event Description

Livanova (b)(4) received a report, that a heater-cooler system 3t was found to be contaminated with mycobacterium chimaera. There is no known patient involvement.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHEATER-COOLER SYSTEM 3T
Type of DeviceCONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80939
GM 80939
Manufacturer Contact
njemile crawley
14401 w. 65th way
arvada, CO 80004
MDR Report Key9522961
MDR Text Key196440643
Report Number9611109-2019-01004
Device Sequence Number1
Product Code DWC
Combination Product (Y/N)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type HEALTH PROFESSIONAL
Reporter Occupation
Type of Report Initial
Report Date 12/27/2019
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received12/27/2019
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number16-02-80
Was Device Available For Evaluation? No Answer Provided
Is The Reporter A Health Professional?
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received11/27/2019
Was Device Evaluated By Manufacturer? No Answer Provided
Is The Device Single Use? No Answer Provided
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015

-
-