ST. JUDE MEDICAL FAST-CATH¿ HEMOSTASIS INTRODUCER, CATH-LOCK¿ LOCKING HUB, 12 CM SHEATH, 5 F; INTRODUCER, CATHETER
|
Back to Search Results |
|
Model Number 406700 |
Device Problem
Material Separation (1562)
|
Patient Problem
Device Embedded In Tissue or Plaque (3165)
|
Event Date 12/10/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission. .
|
|
Event Description
|
Following the procedure, the sheath detached into the patient.While applying pressure near the access site for sheath removal, a portion of the sheath had detached in the leg of the patient.The sheath fragment migrated up to the right atrium where a snare was used to successfully retrieve the fragment.The patient is in stable condition.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Additionally, no lot number was provided so a review of the device history record (dhr) was not possible.Based on the information received, the cause of the reported material fragmentation could not be conclusively determined.
|
|
Event Description
|
Following the procedure, the sheath detached into the patient.While applying pressure near the access site for sheath removal, the sheath had detached near the hub and the shaft detached into the leg of the patient.The sheath fragment migrated up to the right atrium where a snare was used to successfully retrieve the fragment.The patient is in stable condition.
|
|