| Catalog Number |
01.26.48MB |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Batch review performed on 4 december 2019.Lot 165684: (b)(4) items manufactured and released on 25-jan-2017 expiration date: 01/10/2022.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any other similar reported events.Additional implant involved: stem: minimax 01.13.102r cementless anatomical stem right size 2 (k170845) lot.172015.Batch review performed on 4 december 2019 lot 172015: (b)(4) items manufactured and released on 31-aug-2017 expiration date: 18/08/2022.No anomalies found related to the problem.To date, (b)(4) items of the same lot have been already sold without any other similar reported events.Clinical evaluation performed by medical affairs manager hip revision surgery performed 1 year and 8 months after cementless double mobility total hip arthroplasty in a (b)(6) year old woman.Radiographic image provided shows suboptimal cup position.The reason of this choice is unknown: bone morphology or patient anatomy may be one reason but this cannot be confirmed on the basis of a single anteroposterior x-ray.There is no reason to suspect a faulty device.
|
| |
|
Event or Problem Description
|
|
Revision surgery performed 1 year and 9 months after the primary due to impingement (the cup with the stem or the lesser trochanter with the pelvis).The cup was anteverted (the cause is unknown).The surgeon revised the head and liner.Infection not present.
|
| |
|