Model Number 1012451-20
Medical Device Problem Codes Failure to Fold (1255); Difficult to Remove (1528)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/06/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Exemption number e2019001.The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that this was a procedure to treat a lesion with heavy calcification and no tortuosity in the proximal mid left anterior descending (lad) coronary artery.Atherectomy was performed and then a 3.5 x 20 mm voyager nc balloon catheter was advanced to the target lesion without resistance.The balloon was inflated three times to 16 atmospheres.The balloon deflated normally but re-wrap of the balloon was not normal.The balloon deflated like a pancake; therefore, there was difficulty removing the balloon through the guiding catheter.The guiding catheter had to be removed first, then the voyager nc was removed without issue.A stent was implanted successfully completing the procedure.There was no adverse patient effect and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual and functional inspections were performed on the returned device.The reported failure to fold cold not be confirmed.The reported difficulty removing the device from the guiding catheter could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The investigation was unable to determine a conclusive cause for the reported failure to fold however, the reported difficulty removing the device from the guiding catheter appears to be related to operational context.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.D4 - lot no updated, expiration date updated h4 - device manufacture date updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK RX
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9526048
Report Number2024168-2019-15031
Device Sequence Number16836478
Product Code LOX
UDI-Device Identifier08717648151996
UDI-Public08717648151996
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2022
Device Model Number1012451-20
Device Catalogue Number1012451-20
Device Lot Number91015G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Initial Date Received by Manufacturer 12/11/2019
Supplement Date Received by Manufacturer02/08/2020
Initial Report FDA Received Date12/27/2019
Supplement Report FDA Received Date02/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age85 YR
Patient Weight82
-
-