C.R. BARD, INC. (COVINGTON) -1018233 CWS 400 CLOSED WOUND SUCTION KIT, PVC DRAIN; WOUND DRAIN
|
Back to Search Results |
|
Catalog Number 0043610 |
Device Problems
Break (1069); Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 11/27/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
|
|
Event Description
|
It was reported that the wound drain appeared to be pulling mainly air unless the patient was in the supine position.The patient was being helped to get up to the chair, and the patient's dressing was saturated and leaking onto the bed.The incisional dressing pulled back and a small hole was noticed in the drain tubing right by the suture.The drain was pulling blood but instead of going into the collection chamber, it was pooling around the incision.No pulling of tube or odd positioning of tube was noted that may have caused the hole.The hole wasn't a perfect circle, it looked like it was torn even though the tube was sutured in above the hole.No medical intervention was reported.
|
|
Event Description
|
It was reported that the wound drain appeared to be pulling mainly air unless the patient was in the supine position.The patient was being helped to get up to the chair, and the patient's dressing was saturated and leaking onto the bed.The incisional dressing pulled back and a small hole was noticed in the drain tubing right by the suture.The drain was pulling blood but instead of going into the collection chamber, it was pooling around the incision.No pulling of tube or odd positioning of tube was noted that may have caused the hole.The hole wasn't a perfect circle, it looked like it was torn even though the tube was sutured in above the hole.No medical intervention was reported.
|
|
Manufacturer Narrative
|
The reported event was inconclusive as no sample was returned for evaluation.A potential failure mode could be 'loss of vacuum pressure.¿ with a potential root cause of 'tubing/wound drain is damaged.¿ the lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "vii.Instructions for use: 1.The surgeon should irrigate the wound with sterile fluid and then suction the irrigating fluid and gross debris from the operative site.2.Tubes should lie flat and in line with the anticipated skin exit.To facilitate later removal by manual traction, the tubing should not be curled, pinched, or sutured internally.3.Positioning of the drain in the body cavity, as well as the number of drains indicated, should be determined by the surgeon.4.Drain tubing should be placed within the wound by approximating the areas of critical fluid collection.5.Care must be taken to ensure that all drain perforations or channels lie completely within the wound or cavity to be drained.6.Taping or a triple loop suture (around and not through the tubing) will aid in preventing 4 accidental drain placement.7.Deep drainage is best accomplished by using one or more drains for each level of tissue.Each level should be evacuated by a separate vacuum source.8.Care must be exercised to avoid damage to the drain (refer to warnings).The tubing should be repeatedly checked during closure for free motion to avoid breakage and/or fragment retention within the wound.9.When using a trocar please follow these instructions: 9.I.) with one drain: - draw drain using trocar from inside to outside of wound.- ensure that perforated section of the drain is within the critical fluid collection areas of wound.- remove trocar only by cutting the drain one inch from the end of the trocar.- trim non-perforated section of drain to desired length.- attach non-perforated section of drain either to an evacuator inlet port or to a y-connector.9.Ii.) with two single drains: - follow instruction #9.I for each of the two drains separately.9.Iii.) with a double drain: - draw drain using trocar from inside to outside of wound.- ensure that desired perforated region of the drain is within the critical fluid collection areas of wound.- cut the outer portion of the drain (outside the wound area) in the middle of the perforated region.Attach non-perforated section of the inserted drain to an evacuator inlet port or to a y-connector.- after cutting (as mentioned above), the second half of this drain can be used separately.If you are not using the second half then dispose of it as per the hospital protocol.10.Attach drain to evacuator tubing via the y-connector.11.Insert free end of evacuator y-tubing into evacuator suction port a.12.Fully compress evacuator by hand and close drain port b.Unit is now operational.13.To empty unit, clamp y-tubing.Open drain port b.Hold unit with open port at bottom and compress until fluid is removed.14.For continued wound evacuation compress unit fully and close drain port b.Release clamp on y-tubing." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
|