SIEMENS HEALTHCARE DIAGNOSTICS RAPIDPOINT 500 BLOOD GAS ANALYZER; RP 500 Back to Search Results
Catalog Number 10697306
Device Problem Non Reproducible Results (4029)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/05/2019
Event Type  malfunction  
Manufacturer Narrative
The customer stated that repeat testing was performed to confirm correct results which met the clinical picture and a corrected report was issued.Also stated was that in the error logs there were d2 errors for pco2 and "interfering substances detected" messages.Per the rp 500 operator's guide: "d2 excessive drift: pco2": the system turned the parameter off because the sensor exceeded calibration limits".In addition, as per the rp 500 operator's guide: "quaternary ammonium compounds (such as benzalkonium) is a known interfering substance and should be avoided.Review all cleaning solutions to determine if they contain interfering substances.The sensor contacts and the co-ox optics head assembly, which are located behind the measurement cartridge, may be damaged if they get wet.Sensors in the cartridge may be damaged if cleaning solution enters the sample port." the customer stated the measurement cartridge has been changed and they are operational.Data files have been provided for investigation.
 
Event Description
The customer reported discrepant pco2 results on the rp 500 when compared to the results of another rp 500.There was no report of injury due to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS
511 benedict avenue
tarrytown NY 10591
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
northern road
chilton industrial estate
sudbury, CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
6107052212
MDR Report Key9526507
MDR Text Key174176096
Report Number3002637618-2019-00135
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K122539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 12/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10697306
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/11/2019
Initial Date FDA Received12/27/2019
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-