• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE-RITE 8 ULTRASOUND; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770501
Device Problems Loss of Power (1475); Battery Problem (2885)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The device has not been received by the manufacturer, at this time, for evaluation.A history review of serial number (b)(4) showed no other similar product complaint(s) from this serial number.Device has not been received, at this time.
 
Event Description
Per biomed: battery is not holding a charge.Shuts down when not on ac power.
 
Manufacturer Narrative
H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of battery is not holding a charge and shuts down when not on ac power was confirmed.The sr8 was received in good physical condition.The root cause of the reported issue was identified as an internal battery failure.The sr8 was powered on at 88% battery life and ran for around five minutes before shutting down with 86% battery life still on the scanner.The device was serviced, tested, and returned to the customer.A history review of serial number (b)(6) showed no other similar product complaint(s) from this serial number.
 
Event Description
Per biomed - battery is not holding a charge.Shuts down when not on ac power.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE-RITE 8 ULTRASOUND
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key9527040
MDR Text Key178353960
Report Number3006260740-2019-04021
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741098338
UDI-Public(01)00801741098338
Combination Product (y/n)N
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Remedial Action Notification
Type of Report Initial,Followup
Report Date 04/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/27/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770501
Device Catalogue Number9770501
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2020
Event Location Hospital
Date Manufacturer Received04/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number2243072-10/11/2019-013-C
Patient Sequence Number1
-
-