• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL END CAP, STANDARD T2 TIBIA Ø8 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL END CAP, STANDARD T2 TIBIA Ø8 MM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1822-0003S
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Irritation (1941); Reaction (2414)
Date of Event 12/04/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device is not available for return, device is still implanted.
 
Event or Problem Description
It was reported, allergic reaction, skin redness 10 weeks post surgery.
 
Event or Problem Description
It was reported, allergic reaction, skin redness 10 weeks post surgery.
 
Additional Manufacturer Narrative
Received product data revealed the item was manufactured with titanium ti6al4v eli (astm f 136) which is intended for implantation.No indications of material, manufacturing or design related problems were found during the investigation.The reported issue is known within the product's risk management file.Based on the insufficient information provided, the root cause could not be determined.If any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
END CAP, STANDARD T2 TIBIA Ø8 MM
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
DE  D-24232
MDR Report Key9527266
Report Number0009610622-2019-00997
Device Sequence Number10877630
Product Code HSB
UDI-Device Identifier04546540197542
UDI-Public04546540197542
Combination Product (Y/N)N
PMA/510(K) Number
K003018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/29/2024
Device Model Number1822-0003S
Device Catalogue Number18220003S
Device Lot NumberK0F3F32
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/04/2019
Supplement Date Received by Manufacturer01/03/2020
Initial Report FDA Received Date12/27/2019
Supplement Report FDA Received Date01/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age34 YR
-
-