| Model Number |
1320-0090 |
| Medical Device Problem Codes |
Difficult to Remove (1528); Mechanical Jam (2983); Difficult or Delayed Separation (4044)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Once the investigation has been completed any additional information will be reported in a supplemental report.
|
| |
|
Event or Problem Description
|
|
The customer reported to the (b)(6) that : "the incident occurred during the intervention with a gamma nail once the gamma nail was installed, it was impossible to unscrew the nail from the target device.It was necessary to repeat the whole procedure with another instrument panel and once the nail was extracted from the target device, the locking was complete and it was still impossible to unscrew it.The unlocking had to be done in sterilization.The extracted nail has been quarantined with the target device, a replacement of the nail port has been requested".
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event could not be confirmed.Visual inspection: no signs of significant damage and no evidence of rough handling found.A functional inspection showed that the nail could be assembled and disassembled with the nhs to the target device successfully; all products were found fully functional.The alleged event could not be reproduced.A review of the device history for the reported lot did not indicate any abnormalities.A review of the labeling did not indicate any abnormalities.No indications of material, manufacturing or design related problems were found during the investigation.Based on investigation a root cause could not be determined.If any further information is provided, the investigation report will be updated.
|
| |
|
Event or Problem Description
|
|
The customer reported to the ansm that : "the incident occurred during the intervention with a gamma nail once the gamma nail was installed, it was impossible to unscrew the nail from the target device.It was necessary to repeat the whole procedure with another instrument panel and once the nail was extracted from the target device, the locking was complete and it was still impossible to unscrew it.The unlocking had to be done in sterilization.The extracted nail has been quarantined with the target device, a replacement of the nail port has been requested.
|
| |
|