• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC EQUIPMENT RACK 9734060 I7 110V; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC EQUIPMENT RACK 9734060 I7 110V; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 9734060
Medical Device Problem Code Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: other relevant device(s) are: product id: 9735585, system version: 3.1.1 from analysis is not yet complete.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation system.It was reported that the system had a "localizer not connected" error message.The site pressed the niu reset button but this did not resolve the issue.The site rebooted the system and issue resolved.There was a delay to the procedure of less than one hour.There was no impact to the patient related to this event.
 
Additional Manufacturer Narrative
Patient information received, additional information was added to the event description and the initial reporter was received.The manufacturer representative went to the sit to test the navigation system.The reported issue was not confirmed and no parts were replaced.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Additional information was received stating that the issue occurred during a cranial resection and that navigation was not aborted.The issue was noticed in the verify instruments part of the software.
 
Additional Manufacturer Narrative
H3) the software analysis reviewed the logs but provided no additional insight regarding the cause of reported complaint.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUIPMENT RACK 9734060 I7 110V
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9527870
Report Number1723170-2019-06203
Device Sequence Number366485
Product Code HAW
UDI-Device Identifier00613994653482
UDI-Public00613994653482
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K050438
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/30/2012
Device Model Number9734060
Device Catalogue Number9734060
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/26/2019
Supplement Date Received by Manufacturer12/30/2019
01/31/2020
Initial Report FDA Received Date12/27/2019
Supplement Report FDA Received Date01/24/2020
02/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
"SEE H10."
Patient Age42 YR
Patient Weight64
-
-