| Model Number |
9734060 |
| Medical Device Problem Code |
Human-Device Interface Problem (2949)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/24/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: other relevant device(s) are: product id: 9735585, system version: 3.1.1 from analysis is not yet complete.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation system.It was reported that the system had a "localizer not connected" error message.The site pressed the niu reset button but this did not resolve the issue.The site rebooted the system and issue resolved.There was a delay to the procedure of less than one hour.There was no impact to the patient related to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Patient information received, additional information was added to the event description and the initial reporter was received.The manufacturer representative went to the sit to test the navigation system.The reported issue was not confirmed and no parts were replaced.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Additional information was received stating that the issue occurred during a cranial resection and that navigation was not aborted.The issue was noticed in the verify instruments part of the software.
|
| |
|
Additional Manufacturer Narrative
|
|
H3) the software analysis reviewed the logs but provided no additional insight regarding the cause of reported complaint.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|