Medtronic received information that the central part of the tip of the medtronic flow diverter was not visible during the procedure and it was suspected to be damaged.Prior to the event, the medtronic flow diverter was delivered after the medtronic microcatheter was in place and everything was normal before the medtronic flow diverter was opened.After the tip of the medtronic flow diverter was released for the first time, the central part of the tip was not visualized under radiography.The device was retrieved and deployed again, but the results were the same so the medtronic flow diverter was removed from the patient.The patient was undergoing embolization treatment of an unruptured amorphous aneurysm located in the ophthalmic segment of the left internal carotid artery (ica).The patient was on dual antiplatelet therapy (dapt).The distal and proximal landing zone was 3.4mm x 4.3mm.The vasculature was moderate in tortuosity.
|