Model Number CD3231-40Q
Medical Device Problem Codes Premature Discharge of Battery (1057); Difficult to Interrogate (1331)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/09/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
It was reported during an implant of an epicardial lv lead, an attempt to interrogate the device was performed.After multiple attempts and with two different programmers, it was unable to communicate with the device.After discussing the situation, the conclusion was that the battery was early depleted.The device was replaced without problem and the patient was stable after the procedure.
 
Additional Manufacturer Narrative
Premature battery depletion was confirmed by analysis.No sources of high current were noted.The cause of the premature battery depletion was consistent with li cluster formation.From these analyses, in the absence of high current draw, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.Li clusters are a known depletion mechanism for these advisory products that has been investigated and associated with a field action in october 2016.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIFY CRT-D
Common Device Name
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
MDR Report Key9528275
Report Number2938836-2019-17761
Device Sequence Number388811
Product Code NIK
UDI-Device Identifier05414734504393
UDI-Public05414734504393
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P030054
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 04/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2012
Device Model NumberCD3231-40Q
Device Catalogue NumberCD3231-40Q
Device Lot Number3234039
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2019
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/09/2019
Supplement Date Received by Manufacturer04/02/2020
Initial Report FDA Received Date12/28/2019
Supplement Report FDA Received Date04/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0116-2017
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-