Catalog Number 511.606 |
Device Problem
Output below Specifications (3004)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/01/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.However, since the investigation is still on-going, the assignable root cause could not be determined at this time.Once investigation has been completed, a supplemental medwatch will be submitted accordingly.
|
|
Event Description
|
It was reported by(b)(6) that during service and evaluation, it was determined that the air reamer/drill device had low power.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2019.All available information has been disclosed.If additional information should available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Correction: device availability: d10: the device availability date was incorrect in the initial report.The date has been updated from 12/16/2019 to 11/29/2019.Device evaluation: this device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device failed pretest for check power with test stand, minimum 190 w to 260 w (watt).Therefore, the reported condition was confirmed.The assignable root cause was determined to be traced to component failure, which is normal wear (faulty parts).If additional information should become available, a supplemental medwatch report will be submitted accordingly.
|
|