| Model Number |
MMT-523LNAL |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/20/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer reported via phone call that the reservoir lip ring was cracked.The customer¿s blood glucose level was unknown.Customer states the insulin pump has cosmetic damage.The customer also reported that the reservoir lip ring was damaged.The customer reported that the reservoir was unable to lock in place when inserted into insulin pump.The customer was advised to discontinue use of the insulin pump and to revert to the back-up plan.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement.Device received with partial broken reservoir tube lip.The p-cap locks into the reservoir compartment properly noted.
|
| |
|