|
As the reported device was not returned, angiodynamics was unable to perform a device evaluation.The supplier of this device was notified of the event via (b)(4) and requested to perform a device history record review.As per the (b)(4) report, medron, llc, the lot file for 703447 and 76302 was reviewed for any possible causes that could have resulted in the scar.The supplier identified a manufacturing issue as the root cause.Supplier correction/corrective actions: instruction/training.The manufacturing procedure that demonstrates the proper procedure for mini stick sheath molding is being updated with clarifying pictures and instructions that will further demonstrate acceptable and improperly loaded extrusions on core pins.As well as clarifying that this step to verify the extrusion have been loaded properly must be performed when the cassette has been loaded into the mold.All mold operators will be trained to these updates.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: directions for use (dfu).Warning.Contents supplied sterile using an ethylene oxide (eo) process.Do not use if sterile barrier is damaged.If damage is found, call your navilyst medical representative.Inspect prior to use to verify that no damage has occurred in shipping.For single patient use only.Do not reuse, reprocess or resterilize.Reuse, reprocessing or resterilization may compromise the structural integrity of the device and/or lead to device failure which, in turn, may result in patient injury, illness or death.Reuse, reprocessing or resterilization may also create a risk of contamination of the device and/or cause patient infection or cross-infection, including, but not limited to, the transmission of infectious disease(s) from one patient to another.Contamination of the device may lead to injury, illness or death of the patient.After use, dispose of product and packaging in accordance with hospital, administrative and or local government policy.Device description.The mini stick* max coaxial microintroducer kit contains: (1) coaxial sheath/dilator assembly; (1) 21 g (0.9 mm) vascular introducer needle; and (1) 0.018 inch (0.46 mm) floppy tip guidewire.Intended use/ indications for use.The coaxial microintroducer kit is used for the percutaneous introduction of a guidewire into the vascular system.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
|