| Model Number |
H701374 |
| Medical Device Problem Codes |
Display or Visual Feedback Problem (1184); Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/16/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission. .
|
| |
|
Event or Problem Description
|
|
During a cryo atrial fibrillation procedure, a rattling noise was heard within the cim and the unit displayed dark images so the procedure was cancelled.There were no adverse consequences to the patient due to the cancellation.
|
| |
|
Additional Manufacturer Narrative
|
|
The results of the investigation are inconclusive since the device was not returned for analysis.The cim was replaced in the field and the customer retained the unit so it will not be returned.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information received, the cause of the reported display issue and subsequent procedure cancellation could not be conclusively determined.
|
| |
|