| Model Number |
393009 |
| Medical Device Problem Code |
Premature Elective Replacement Indicator (1483)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
This device was explanted and replaced due to eri.The device was programmed at max output for over a year due to a myopore lv lead with high thresholds - 7.5v at 1.5ms at the time of explant.Lead had steadily increasing thresholds over the last year.Lv lead was explanted and replaced as well.No adverse patient events were reported.Should additional information become available, this file will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
We received your event description for the above mentioned device and would like to thank you for supporting our post-market surveillance.The device is currently not available for analysis.Therefore, no conclusion can be drawn at this time.However, biotronik monitors the post-market performance of its products closely in order to identify any trends that may reveal over time a potential manufacturing or design issue.The historic performance of the product involved in the current case does not at this point reveal any such trend.Should additional relevant information become available, this investigation will be updated.
|
| |
|