Model Number 232453
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 12/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
It was reported by the sales rep via phone that during an acl procedure the rigidloop combo reamer/passing pin 4.5mm broke and they were able to remove the fragments from the patient´s body.The procedure was successfully completed.2 minutes delay was reported.No patient consequences.Additional information received from the affiliate stated the following: "it could be possible that the beath pin hit the ¿over the top guide¿ as it began reaming through the femur which caused the tip to break.However, i was paying close attention to this and did not notice that scenario happening.Other than that i can't imagine what would've caused the breakage other than the force of the pin being drilled through the femur.".
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.One possible root cause of the reported problem could be the beath pin hit the 'over the top guide' as it began reaming through the femur which caused the tip to break.However without the return of the complaint device, and no further information provided we cannot determine a definitive root cause.A manufacturing record evaluation was performed for the finished device lot number (2l25117), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot
=
>a manufacturing record evaluation was performed for the finished device lot number (2l25117), and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMBO BEATH PIN EA
Common Device Name
CANNULATED SURGICAL DRILL
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9531487
Report Number1221934-2019-60200
Device Sequence Number11341510
Product Code HWE
UDI-Device Identifier10886705024544
UDI-Public10886705024544
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model Number232453
Device Catalogue Number232453
Device Lot Number2L25117
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/16/2019
Supplement Date Received by Manufacturer01/27/2020
Initial Report FDA Received Date12/30/2019
Supplement Report FDA Received Date01/29/2020
Patient Sequence Number1
-
-