• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION MAXPLUS EXTENSION SET; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION MAXPLUS EXTENSION SET; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number MPX9103
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Information (3190)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product will be returned per customer.The customer complaint could not be confirmed because the product was discarded and not returned for failure investigation.The root cause of this failure was not identified.
 
Event or Problem Description
It was reported that a extension set tubing separated when attached to the patient at the point of connection between the y-site port and the male luer.It was still inserted into the patient¿s vein.
 
Event or Problem Description
It was reported that an extension set tubing separated when attached to the patient, at the point of connection between the y-site port and the male luer.It was still inserted in the patient¿s vein.Although requested, there has been no further patient or event information made available to date.
 
Additional Manufacturer Narrative
Correction; h.6.(patient code).The customer complaint of maxguard extension set tubing broken was confirmed.A photo provided by the customer shows that the non-dhep tubing was separated from the outlet port.Previous investigations have concluded that the root cause of tubing separation issues complaints have concluded to be due to insufficient solvent being applied at the affected engagement due to equipment and/ or operator error.No product will be returned per customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAXPLUS EXTENSION SET
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key9531633
Report Number9616066-2019-03764
Device Sequence Number14113811
Product Code FPA
UDI-Device Identifier10885403237294
UDI-Public10885403237294
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K051499
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/01/2024
Device Model NumberMPX9103
Device Catalogue NumberMPX9103
Device Lot Number19105069
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/30/2019
Supplement Report FDA Received Date02/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-