MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX3 PEEK ANC.W/OCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX3 PEEK ANC.W/OCORD; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 222207
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); Foreign Body In Patient (2687)
Date of Event 12/16/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
It was reported by the sales rep via phone that during a rotator cuff repair procedure the tip of their 5.5 healix 3 peek anchor with orthocord broke during insertion.The tip was pulled out and debris was left in the patient.The procedure was completed with another like device using the same bone hole with no patient harm or surgical delay to the case.The patient's bone quality was average.The device was discarded by the customer.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H6: method codes: device history lot: a manufacturing record evaluation was performed for the finished device lot number [4l24048], and no non-conformances were identified.Investigation summary: the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.This complaint cannot be confirmed.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number [4l24048], and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX3 PEEK ANC.W/OCORD
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9531927
Report Number1221934-2019-60203
Device Sequence Number11341198
Product Code MAI
UDI-Device Identifier10886705007554
UDI-Public10886705007554
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K150209
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Model Number222207
Device Catalogue Number222207
Device Lot Number4L24048
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 12/16/2019
Supplement Date Received by Manufacturer02/03/2020
Initial Report FDA Received Date12/30/2019
Supplement Report FDA Received Date02/04/2020
Patient Sequence Number1
-
-