| Model Number |
100143451 |
| Medical Device Problem Code |
Sparking (2595)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
12/13/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
| |
|
Event or Problem Description
|
|
During set up for a procedure the ensite precision link power supply cable fell apart and sparked.A different power supply was utilized for the scheduled procedure.There were no user consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
One ensite precision¿ power supply unit 4 was received.Visual inspection of the returned power supply module confirmed the reported spark as the four pin redel connector has been separated from the outer shell.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The review determined the process was performed and completed in accordance with abbott specifications and procedures.Based on the information provided to abbott and the investigation performed, the root cause of the reported spark was isolated to a detached redel connector body.
|
| |
|