|
STRYKER ORTHOPAEDICS-MAHWAH V40 COCR LFIT HEAD 44MM/+4; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
|
Back to Search Results |
|
| Catalog Number |
6260-9-244 |
| Medical Device Problem Code |
Corroded (1131)
|
| Health Effect - Clinical Codes |
Pain (1994); Injury (2348); Reaction (2414)
|
| Date of Event |
12/03/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Although the lot code was not provided, the event details, catalog number and date of implant are all in scope of ra 2016-028 therefore, this device is being assessed as under the scope of the recall.Conclusion: lot specific voluntary recall was initiated for the lfit v40 cocr heads within scope of a capa.The investigation revealed that only specific catalog numbers and specific lots are impacted by the regulatory action and that the affected lots were manufactured on or before march 4, 2011.The root cause analysis identified a process related anomaly as to the affected sizes and lots.The affected product has all been implanted and/or expired.Device evaluated by manufacturer? not returned to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient's right hip was revised.After patient complaint of pain, diagnostics revealed elevated serum ion levels in the patient's blood.Intra-operatively, some corrosion was noted inside the femoral head.The femoral head was revised to a sleeve with ceramic head.The stem was not revised as the trunnion was intact.Rep reported that no further information will be released by the hospital or surgeon.
|
| |
|
Search Alerts/Recalls
|
|
|