• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. EVOS VOLAR PLATE 4H LEFT STD STD TI 56MM; NAIL, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. EVOS VOLAR PLATE 4H LEFT STD STD TI 56MM; NAIL, FIXATION, BONE Back to Search Results
Model Number 71158104
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/10/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery an evos wrist ti 4h std l volar distal radius plate (71158104) was used with a 16mm locking screw (74412416), but the screw fell through the plate.Then, an evos wrist ti 3h std l volar distal radius (plate 71158103) with a new 16mm screw (74412416) was tried and it had the same issue.The surgery was completed using an old-style volar distal radius set.Additionally, 11mm cortex screw (74402411) was found to have the same failure as well.For this event, there was a delay greater than 30 min and the patient was not injured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS VOLAR PLATE 4H LEFT STD STD TI 56MM
Common Device Name
NAIL, FIXATION, BONE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9532182
Report Number1020279-2019-04671
Device Sequence Number11364743
Product Code JDS
UDI-Device Identifier00885556693827
UDI-Public00885556693827
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71158104
Device Catalogue Number71158104
Device Lot Number19FT23429
Initial Date Received by Manufacturer 12/10/2019
Supplement Date Received by Manufacturer12/16/2020
Initial Report FDA Received Date12/30/2019
Supplement Report FDA Received Date12/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
-
-