• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. PARTIAL ARTICULAR SURFACE LEFT MEDIAL SIZE F 11 MM THICKNESS; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. PARTIAL ARTICULAR SURFACE LEFT MEDIAL SIZE F 11 MM THICKNESS; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 11/22/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2019-05607 and 0001825034-2019-05787.Concomitant medical products: unknown persona partial knee femoral: catalog#: ni, lot#: ni.Unknown persona partial knee tibial tray: catalog#: ni, lot#: ni.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that during a knee procedure, the tibial insert would not seat on the tibial tray.The procedure was completed using another tibial insert.No adverse events have been reported as a result of the malfunction.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Visual inspection of the returned articular surface exhibits signs of use (nicks and gouges) and the device exhibits damage and is compressed at one side caused during attempts to insert the prosthesis.Dimensional measurements could not be taken because of the damage.Device history record was reviewed and no discrepancies were found.The compressed feature of the insert indicates that the device was not properly inserted.The root cause is attributed to use error.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARTIAL ARTICULAR SURFACE LEFT MEDIAL SIZE F 11 MM THICKNESS
Common Device Name
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9532285
Report Number0001825034-2019-05607
Device Sequence Number2313550
Product Code HSX
UDI-Device Identifier00880304818323
UDI-Public(01)00880304818323(17)231031(10)64157510
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
PMA/510(K) Number
K161592
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberN/A
Device Catalogue Number42518200611
Device Lot Number64157510
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/30/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 12/03/2019
Supplement Date Received by Manufacturer03/17/2020
Initial Report FDA Received Date12/30/2019
Supplement Report FDA Received Date03/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Patient Age82 YR
-
-