It was reported by the rn that the intra-aortic balloon pump (iabp) alarmed for purge failure and system error 7.
The rn reported that the patient was transferred from an outside facility with a competitors iab.
The clinical support specialist (css) walked the rn through the connections to the pump and therapy started.
The rn reported that the screen "wasn't acting right".
The css facetime with the staff, it was noted that the touchscreen needed a calibration when they attempted to pull up the arterial pressure (ap) window to zero.
When the staff begin touchscreen calibration, a system error 7 was issued.
The alarm was silenced by pressing the corner switch several times.
Once the touchscreen calibration came up, the first target wouldn't respond.
The css then had the staff power down and back up and the pump would restart therapy.
The pump was swapped out.
Pumping was interrupted for 1 minute.
There was no report of patient injury or consequence.
There were no complications, serious injury or death reported.
|