| Model Number |
FB-211K |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/11/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the service center for evaluation.Therefore, the exact cause of the reported event cannot be determined.Upon receipt of the complaint device, an evaluation will be performed and a report will be submitted.The instruction manual contains several warning statements in an effort to prevent damage to the biopsy forceps; "when inserting the instrument into the endoscope, hold the slider firmly.Otherwise, the cups may open and extend from the distal end of the endoscope abruptly which may cause damage to the endoscope and /or instrument.Do not insert the instrument into the endoscope while the cups are open.It could also damage the endoscope and/ or instrument.If you use the instrument several times during one procedure, confirm that there is no irregularity with its operation and appearance, each time you use it.Do not use the instrument if you detect any abnormality.The breakage of the distal end such as the manipulation wire's detachment could cause mucous membrane damage.".
|
| |
|
Event or Problem Description
|
|
The service center received a report of biopsy forceps (fb-211k) that had one side of the jaw fall off and into the patient during an unspecified procedure.The physician was able to retrieve the detached piece.It was reported there was no patient injury but it is unknown if the intended procedure was completed.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the additional information received from the reporting facility.Please see the updated information in sections: a2, a3, a4, b5, b7, e2, e3, g3, g4, g7, h2 and h10.
|
| |
|
Event or Problem Description
|
|
The service center received a report of a biopsy forceps (fb-211k) that had a jaw break off during a procedure, and fell into the patient.The fb-211k was inspected prior to the procedure and no anomalies were observed.The physician was in the middle of an endoscopy with biopsy procedure, when the jaw of the forceps broke off and fell into the esophagus of the patient.The item was retrieved by the physician using a steris roth net.It was reported this event delayed the procedure approximately for five minutes and there was no blood loss to the patient.The report indicated no other instruments were replaced during the procedure, and that a second set of forceps was used to continue taking esophageal biopsies.There was no patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the device evaluation and correction of the lot number and manufacture date.The device was returned to the service center for evaluation.The received condition device for the reported complaint of ¿jaw of forceps fell off¿ was confirmed.A visual inspection was performed on the received device and there was damage observed on the forceps.The distal end of the forceps was observed to have one side of the alligator jaw broken.The broken jaw was not returned to the service center for evaluation.The insertion portion of the device was inspected and noted to have two kinks in the middle section and at the distal end section.The slider was tested and observed to operate normally and as intended.Based on the evaluation of the received condition disposable biopsy forceps (fb-211k), lot number 92v 14, the reported event was confirmed, and determined to be caused by the excessive force, which was attributed to mishandling of the device.
|
| |
|
Search Alerts/Recalls
|