• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER POLYFLUX 17L; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER POLYFLUX 17L; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Catalog Number 109650
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 11/26/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that before use patient use of a polyflux 17l apac, a particulate matter was observed inside the cap.There was no patient involvement reported.No additional information is available.
 
Manufacturer Narrative
Additional information added to d10, h3, h6 and h10.H10: the device was received for evaluation.During visual inspection by naked eye of the product, a hair in the blood side was found.The hair was clamped between the o-ring and the polyurethane cutting surface.The reported condition was verified.The cause could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.A nonconformance has been opened to address this issue.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYFLUX 17L
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
deerfield IL
MDR Report Key9533657
MDR Text Key185674018
Report Number9611369-2019-00198
Device Sequence Number1
Product Code KDI
UDI-Device Identifier07332414086626
UDI-Public(01)07332414086626
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/30/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2022
Device Catalogue Number109650
Device Lot Number9-4210-H-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/23/2019
Date Manufacturer Received01/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-