| Model Number |
G21396 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
12/06/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is still pending, a follow up mdr will be submitted to include the investigation conclusions.
|
| |
|
Event or Problem Description
|
|
The device was used for erbd with the approach from the duodenal papilla.After placing a wire guide (visiguide 0.025inch/olympus) at the common bile duct, the physician attempted to straighten the complaint device with the pigtail straightener and to made it follow the wire guide.However, he could not do that because the pig tail part got kinked.Therefore, the procedure was completed with zebd-7-4.Event is fda mdr reportable based on the device malfunction reporting precedence for this device family for the issue of ¿stent kinked/ bent'.
|
| |
|
Event or Problem Description
|
|
The device was used for erbd with the approach from the duodenal papilla.After placing a wire guide (visiguide 0.025 inch/olympus) at the common bile duct, the physician attempted to straighten the complaint device with the pigtail straightener and to made it follow the wire guide.However, he could not do that because the pig tail part got kinked.Therefore, the procedure was completed with zebd-7-4.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: 1 x zso-7-7 of lot # c1656543 was returned to cirl for evaluation open in its original packaging.Lab evaluation: the device involved in the complaint was evaluated in the laboratory on 10th january 2020.The returned device lab findings and observations can be referred through the attached files.A kink was noted on the 5th porthole on the pigtail.A 0.035¿ wire guide will not pass the kink.Note: no straightener was returned with the device image review: n/a.Document review including ifu review: prior to distribution zso-7-7 devices are subjected to a visual inspection and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for zebd-7-7 of lot number c1656543 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1656543.It should be noted that the instructions for use (ifu0045-7) states the following: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use.¿ ¿care must be exercised when straightening the pigtail curls in order to avoid kinking or breaking the stent.¿ root cause review: a definitive root cause can be attributed to the use of a non-recommended wire guide with the device.It may be noted the device label indicates a 0.035¿ wire guide for use with this device and all simulated use testing to date has been completed using a 0.035¿ wire guide.As the smaller diameter 0.025¿ wire guide occupies less space in the lumen of the pigtail stent it is likely that the extent of the curvature of the pigtail will be greater when a 0.025¿ wire guide is used.As per additional information received a 0.025" wire guide was used.Summary: complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.Complaints of this nature will continue to be monitored for potential emerging trends.
|
| |
|