Model Number G17147
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code Discomfort (2330)
Date of Event 12/13/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Occupation: sterile technician.This report includes information known at this time.A follow up report will be submitted should additional relevant information become available.
 
Event or Problem Description
It was reported, the doctor had completed a sonohysterogram using a cook silicone balloon hysterosalpinography injection catheter, and when he attempted to deflate the device he found that the balloon tip would not deflate.The doctor had to pull the catheter out fully inflated.No unintended section of the device remained inside the patient.The patient was reported to be "fine", she just experienced discomfort during the procedure.
 
Event or Problem Description
No new patient or event information since the last report was submitted.
 
Additional Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unchanged, or unavailable.Ec method code desc - 5: communication/interviews (4111).Investigation ¿ evaluation: a visual inspection and functional testing of the returned device was conducted.A document based investigation was also performed including a review of complaint history, device history record, the instructions for use, manufacturing instructions, and quality control data.One device was returned for investigation.Visual examination of the returned device noted the catheter was bent 1.5cm and 19.5cm from the distal tip the balloon deflated, but pressure was needed when depressing the plunger.Functional testing determined the balloon inflated properly.A review of the device history record found no non-conformances related to the reported failure mode.A review of complaint history records shows one other complaint associated with the complaint device lot.Due to the individual nature of the manufacturing and inspection process for the devices in the lots, it is unlikely that these events are an indication of device issue within the entire lot.The instructions for use (ifu), provides the following information to the user related to the reported failure mode: warnings "always inflate the balloon with a sterile liquid.Never inflate with air, carbon dioxide or any other gas." instructions for using cook silicone balloon hsg catheter saline infusion sonohysterography 6.Fill the syringe with sterile saline solution.Eliminate air from the syringe.7.To inflate the balloon, instill sterile saline solution." how supplied "upon removal from the package, inspect the product to ensure no damage has occurred." a review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was visually/functionally inspected by quality control and no related gaps in production or processing controls were noted.Cook has concluded the operator's failure to follow instructions contributed to this incident.Per the quality engineering risk assessment, no further action is warranted.Cook medical will continue to monitor this device via the complaints database for similar complaints.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COOK SILICONE BALLOON HYSTEROSALPINGOGRAPHY INJECTION CATHETER
Common Device Name
HES INSUFFLATOR-CARBON DIOXIDE, URETOTUBAL (AND ACCESSORIES)
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key9535725
Report Number1820334-2019-03167
Device Sequence Number11597560
Product Code HES
UDI-Device Identifier00827002171473
UDI-Public(01)00827002171473(17)211130(10)9347392
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K891290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2021
Device Model NumberG17147
Device Catalogue NumberJ-CHSG-503000
Device Lot Number9347392
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/15/2020
Initial Date Received by Manufacturer 12/13/2019
Supplement Date Received by Manufacturer01/29/2020
Initial Report FDA Received Date12/31/2019
Supplement Report FDA Received Date01/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-