Model Number 1040007
Device Problem Failure to Recalibrate (1517)
Patient Problem No Patient Involvement (2645)
Event Date 12/04/2019
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator would not pass service testing.The device was not in patient use.The ventilator was returned to the manufacturer for evaluation and the customer's complaint was confirmed.The device's oxygen blending module flow sensor was cleaned to address the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRILOGY 202
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
adam price
312 alvin drive
new kensington, PA 15068
7243349303
MDR Report Key9536225
MDR Text Key173335059
Report Number2518422-2019-02992
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093905
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1040007
Device Catalogue Number1040007
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2019
Date Manufacturer Received12/04/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/12/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-