COOK INC PERFORMER INTRODUCER; DYB INTRODUCER, CATHETER Back to Search Results
Catalog Number RCFW-12.0-38-30-RB
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Pma/510(k) number = pre-amendment.Investigation-evaluation: a review of the drawing, instructions for use (ifu), manufacturing instructions, and quality control procedures was conducted during the investigation.The complaint device was not returned; therefore, no physical examinations could be performed; however, a document-based investigation evaluation was performed.There is no evidence to suggest the product was not made to specifications.The lot number of the device is not known; accordingly, a review of the device history record could not be conducted.The device is shipped with instruction for use (ifu) which states: "all instruments or catheters used with this product should move freely through the valve and sheath.Damage to the valve/introducer may result when the fit is tight.When inserting, manipulating, or withdrawing a device through an introducer always maintain introducer position.Do not attempt to insert or withdraw the wire guide and/or introducer if resistance is felt.Sheath introduction: upon removal from package, ensure the inner diameter (id) of the introducer is appropriate for the maximum diameter of the instrument or catheter to be introduced.Using the side-arm of the valve, flush the introducer by filling the introducer assembly completely with heparinized saline.Flush the dilator with heparinized solution.Insert the dilator completely into the introducer.Using standard seldinger technique, access the target vessel with the appropriate needle.Insert wire into the vessel through the needle, then remove needle, leaving the wire guide in place.Insert dilator/sheath combination over wire guide.Remove wire guide and dilator, aspirate and flush introducer side-arm.Insert appropriately sized device as needed.Sheath removal: insert wire guide at least 10cm past the tip of the sheath.Insert the dilator over the wire into the sheath.Withdraw the sheath and dilator as a unit.Remove the wire guide." based on the information provided, no product returned and the results of our investigation, a definitive root cause could not be established.We will continue our monitoring of similar complaints and have notified the appropriate personnel of this event.Per the quality engineering risk assessment, no further action is required.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.
 
Event Description
As reported, during an endovascular aneurysm repair, a performer introducer leaked at the hub.The user reported that the silicone disk does not adjust to wires "like before" and when unspecified lunderquist wires are left inside the device, it leaks blood "drop by drop".A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERFORMER INTRODUCER
Type of Device
DYB INTRODUCER, CATHETER
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
larry pool
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key9536268
MDR Text Key176050032
Report Number1820334-2019-03186
Device Sequence Number1
Product Code DYB
Combination Product (y/n)N
Reporter Country CodeES
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 12/31/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received12/31/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberRCFW-12.0-38-30-RB
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/20/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-