MAUDE Adverse Event Report: MDT SOFAMOR DANEK PUERTO RICO MFG DIVERGENCE ANTERIOR CERVICAL FUSION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT SOFAMOR DANEK PUERTO RICO MFG DIVERGENCE ANTERIOR CERVICAL FUSION SYSTEM Back to Search Results
Catalog Number G6626747
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/04/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that intra-op, the implant cracked near screw holes during insertion.The cracked implant was then completely explanted; and was replaced with a new one.No patient complications were reported due to this event.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient had a pre-operative diagnosis of degenerative disk disease; and underwent anterior cervical discectomy and fusion (acdf) at c3-c4.
 
Additional Manufacturer Narrative
Product analysis: visual and optical inspection revealed the implant has been severely damaged.The screw insertion holes have been collapsed and cracked.The lock nut is no longer attached to the implant.The implant appears to be damaged due to material overload during the insertion process.Unable to determine root cause of damage.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIVERGENCE ANTERIOR CERVICAL FUSION SYSTEM
Manufacturer (Section D)
MDT SOFAMOR DANEK PUERTO RICO MFG
barrio marianna rd 909, km0.4
humacao PR 00792
MDR Report Key9538630
Report Number1030489-2020-00004
Device Sequence Number16804743
Product Code OVE
UDI-Device Identifier00643169337411
UDI-Public00643169337411
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K141599
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 06/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberG6626747
Device Lot NumberH5564369
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/26/2020
Initial Date Received by Manufacturer 12/04/2019
Supplement Date Received by Manufacturer01/02/2020
05/04/2020
Initial Report FDA Received Date01/02/2020
Supplement Report FDA Received Date01/27/2020
06/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-