NELLCOR PURITAN BENNETT MEXICO SA NELLCOR; OXIMETER Back to Search Results
Model Number D-YS
Medical Device Problem Code Low Readings (2460)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, during use, the sensor gave an error message and only showed spo2 reading of 80% while with a new sensor, it showed 100% spo2 reading.There was no patient injury.
 
Additional Manufacturer Narrative
H3 evaluation summary: one device was received for evaluation and the reported issue was confirmed.The emitter¿s internal black and red wires were broken close to the sensor.This type of fault is detected in our manufacturing process for all sensors (disposables or reusable) using two different stations.Product does not leave the manufacturing plant in the condition found in the sample.This condition could be caused by normal usage/bending of the sensor.According to the product risk assessment for device sensors, no mitigation for damaged part (broken wires) required; however design verification 100% in manufacturing testing and residual risk is acceptable.Also dfmea for device sensors, indicates no mitigation required for ¿sensor cable conductor fails open (broken wire)¿ and residual risk also is acceptable.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX  22225
MDR Report Key9538825
Report Number2936999-2020-00010
Device Sequence Number19427085
Product Code DQA
UDI-Device Identifier20884522042540
UDI-Public20884522042540
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K012891
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberD-YS
Device Catalogue NumberD-YS
Device Lot Number190880087H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2020
Initial Date Received by Manufacturer 12/19/2019
Supplement Date Received by Manufacturer03/11/2020
Initial Report FDA Received Date01/02/2020
Supplement Report FDA Received Date04/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-