According to the reporter, during use, the sensor gave an error message and only showed spo2 reading of 80% while with a new sensor, it showed 100% spo2 reading.There was no patient injury.
|
|
H3 evaluation summary: one device was received for evaluation and the reported issue was confirmed.The emitter¿s internal black and red wires were broken close to the sensor.This type of fault is detected in our manufacturing process for all sensors (disposables or reusable) using two different stations.Product does not leave the manufacturing plant in the condition found in the sample.This condition could be caused by normal usage/bending of the sensor.According to the product risk assessment for device sensors, no mitigation for damaged part (broken wires) required; however design verification 100% in manufacturing testing and residual risk is acceptable.Also dfmea for device sensors, indicates no mitigation required for ¿sensor cable conductor fails open (broken wire)¿ and residual risk also is acceptable.Information has been added to the database and trends will continue to be monitored.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|