• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY FRANCE SAS - 3003895575 XTND GLENO D38MM +4MM; SHOULDER IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY FRANCE SAS - 3003895575 XTND GLENO D38MM +4MM; SHOULDER IMPLANT Back to Search Results
Model Number 1307-64-138
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code Not Applicable (3189)
Date of Event 12/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that glenosphere threads would not engage into the metaglene.Waste 3 glenosphere until the metaglene was revised.Threads appeared to be damaged.Surgeon suggested that perhaps the threads in the metaglene were not spec.30 minutes of surgical delay.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: evaluation of the returned device confirms the reported event.This analysis does not highlighted any supplier defect and the need of corrective actions is not indicated.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XTND GLENO D38MM +4MM
Common Device Name
SHOULDER IMPLANT
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key9538901
Report Number1818910-2020-00109
Device Sequence Number19439657
Product Code KWS
UDI-Device Identifier10603295502456
UDI-Public10603295502456
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K183077
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1307-64-138
Device Catalogue Number130764138
Device Lot NumberD19031701
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/17/2019
Initial Date Received by Manufacturer 12/09/2019
Supplement Date Received by Manufacturer02/13/2020
Initial Report FDA Received Date01/02/2020
Supplement Report FDA Received Date03/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age80 YR
-
-