| Model Number |
7055M-90 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
12/09/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmea, the most probable root cause is user error: improper implant selection, using too long of an implant or installing the implant to deep.
|
| |
|
Event or Problem Description
|
|
The patient had left si joint arthrodesis in (b)(6) 2019 where three implants were installed.The patient later reported left side radicular pain.The surgeon determined that the superior positioned implant was impinging on the neuroforamen.In (b)(6) 2019, the surgeon performed a revision procedure where he removed the superior positioned implant.No other preexisting implants were adjusted or removed.The status of the patient following the revision procedure is not known.
|
| |
|