• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDIVANCE, INC. ¿ 1725056 ARCTICSUN 5000 ARCTIC SUN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDIVANCE, INC. ¿ 1725056 ARCTICSUN 5000 ARCTIC SUN Back to Search Results
Model Number 50000000E
Device Problems Use of Device Problem (1670); Therapeutic or Diagnostic Output Failure (3023)
Patient Problems No Consequences Or Impact To Patient (2199); Inadequate Pain Relief (2388)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete, a supplemental report will be filed.
 
Event Description
It was reported that the arctic sun device did not warm a patient. Since the device had cooled to 10c, ms&s advised the biomed to drain 500mls of water and set the water temperature to 40c. The biomed drained 1000mls of water and did not test it. Ms&s advised the biomed to do a total drain and to add 3. 5liters of sterile water with a bottle of cleaning solution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTICSUN 5000
Type of DeviceARCTIC SUN
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
louisville CO 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington, GA 30014
7707846100
MDR Report Key9540051
MDR Text Key178033016
Report Number1018233-2020-00028
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 02/18/2020
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received01/02/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number50000000E
Device Catalogue Number50000000E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received02/13/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

-
-