• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. VIEWFLEX¿ PLUS ICE CATHETER INTERFACE MODULE FOR VIEWMATE Z INTRACARDIAC ULTRASO; ULTRASONIC PULSED DOPPLER IMAGING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. VIEWFLEX¿ PLUS ICE CATHETER INTERFACE MODULE FOR VIEWMATE Z INTRACARDIAC ULTRASO; ULTRASONIC PULSED DOPPLER IMAGING SYSTEM Back to Search Results
Model Number 100043720
Medical Device Problem Code Arcing (2583)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/24/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
One viewflex¿ plus ice catheter interface module for viewmate z intracardiac ultrasound console was received for investigation.Trace amounts of foreign material was identified on the transducer connector port.Disassembly was performed, and heavy amounts of saline deposits were identified on numerous internal assemblies with signs of arcing between electronic components.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the information provided to abbott and the investigation performed, there was fluid ingress to the internal assemblies resulting in hardware damage.
 
Event or Problem Description
This event is being reported based on the analysis of the returned device. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIEWFLEX¿ PLUS ICE CATHETER INTERFACE MODULE FOR VIEWMATE Z INTRACARDIAC ULTRASO
Common Device Name
ULTRASONIC PULSED DOPPLER IMAGING SYSTEM
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
one st. jude medical drive
st. paul MN 55117
MDR Report Key9540330
Report Number2184149-2019-00259
Device Sequence Number2664125
Product Code IYN
UDI-Device Identifier05415067002907
UDI-Public05415067002907
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K120703
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 01/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number100043720
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2019
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/11/2019
Initial Report FDA Received Date01/02/2020
Patient Sequence Number1
Patient Age37 YR
Patient Weight96
-
-