| Model Number |
1650DE |
| Medical Device Problem Code |
Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
11/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.The battery was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.A review of the manufacturing documentation confirmed that the associated device met all requirements for release.Failure analysis of the returned device revealed that the device passed visual examination.Functional testing revealed a cell-under-voltage flag and a pack-under-voltage flag, which rendered the battery inoperable.A cell-under-voltage flag is triggered when a cell pair falls below a voltage threshold.Likewise, a pack-under-voltage flag is triggered when the battery pack falls below a voltage threshold.Internal inspection of the battery revealed a defective solder connection between cell pairs of the battery, which contributed to the flags.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to a defective solder.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
The battery was returned to manufacturer because the battery was not charging.The battery subsequently tested out of specification during manufacturer's analysis.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction h10: this report contains an update to the product event summary to reference the formal investigation associated with the reported failure.The previously reported failure and investigation findings remain unchanged.Manufacturer name and contact details updated.Product event summary: an internal investigation evaluated batteries with defective solders.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|