|
Additional information - this emdr is being submitted to include the below: d4: unique identifier (udi) h6: investigation results under imdrf cause investigation code, imdrf investigation findings, imdrf cause conclusions h10: investigation summary an investigation was completed for this complaint by the assigned manufacturing site investigation team.The complaint summary of this investigation required rework.A corrective actions/preventive actions (capa) was raised and an updated/corrected investigation summary has been completed in accordance with assigned capa.A supplemental mdr is being submitted to document the completion of investigation rework and final investigation; the complaint record will proceed to closure.Batch record review: lot 8g04167 was manufactured on 07/01/2018, line form fill and seal #1.With a total of (b)(4) market units.Complaint investigator performed a batch record review on 08/05/2020, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, under international commodity code (icc) code 421035.System application product (sap) material 1709724 and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.There is a photograph associated with this case and no unused return sample was expected.Investigation conclusion: based on the investigation findings, the root cause identified for the issue ¿centricity issues¿, reported under failure mode ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur¿ is attributed to: 1)machine: the investigative process concluded that all the machinery/ tooling items used in the manufacturing process complied when compared against and procedure instruction (pi) pi31-087 ver.10.0 specifications; as however, as part of this investigation it was identified that some size did not fit correctly in the disc assembly station produce the concentricity issues in the accordion line, nevertheless the pi1-087 procedure was recently updated to version:9.0 to modify the ((b)(4)) tooling and a new (b)(4) tooling was added to avoid problems of concentricity on the disks, due to these tooling shared disk size and at the time of assembly some size did not fit correctly in the welding process and caused concentricity issues during the manufacturing process.After this updated was performed the manufacturing line demonstrated to behave steadily after the correction.No issues were identified for material, measurement, method and environment investigations.The investigation associated with related event has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
|