• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CONVATEC DOMINICAN REPUBLIC INC ACCORDION FLANGE; PROTECTOR, OSTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONVATEC DOMINICAN REPUBLIC INC ACCORDION FLANGE; PROTECTOR, OSTOMY Back to Search Results
Model Number 421035
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Device 1 of 30.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
 
Event or Problem Description
The nurse reported that the wafer had an off-centered starter hole.The product was not used.Picture of the alleged malfunction was provided by the complainant.
 
Additional Manufacturer Narrative
H10: investigation summary: an investigation was completed for this complaint by the assigned manufacturing site investigation team.The complaint summary of this investigation required rework.A corrective actions/preventive actions (capa) was raised and an updated/corrected investigation summary has been completed in accordance with assigned capa.A supplemental mdr is being submitted to document the completion of investigation rework and final investigation; the complaint record will proceed to closure.Batch record review: lot 8g04167 was manufactured on 07/01/2018, line form fill and seal #1.With a total of (b)(4)market units.Complaint investigator performed a batch record review on 08/05/2020, to verify if all the applicable procedures were followed and no issues were found; all the components for assembly were correct per bill of materials (bom) and all the tooling information documented was also correct, under international commodity code (icc) code 421035.System application product (sap) material 1709724 and manufacturing order (b)(4).The batch record review supports that there were no discrepancies related to the issue reported.There is a photograph associated with this case and no unused return sample was expected.Investigation conclusion: based on the investigation findings, the root cause identified for the issue ¿centricity issues¿, reported under failure mode ¿skin barrier starter hole is defective (e.G misalignment or off center), leakage may occur¿ is attributed to: 1)machine: the investigative process concluded that all the machinery/ tooling items used in the manufacturing process complied when compared against and procedure instruction (pi) pi31-087 ver.10.0 specifications; as however, as part of this investigation it was identified that some size did not fit correctly in the disc assembly station produce the concentricity issues in the accordion line, nevertheless the pi1-087 procedure was recently updated to version:9.0 to modify the (6qnf0049 and 6qnf0050) tooling and a new 6qnf0004 tooling was added to avoid problems of concentricity on the disks, due to these tooling shared disk size and at the time of assembly some size did not fit correctly in the welding process and caused concentricity issues during the manufacturing process.After this updated was performed the manufacturing line demonstrated to behave steadily after the correction.No issues were identified for material, measurement, method and environment investigations.The investigation associated with related event has been approved and is complete.No additional action is required, and this complaint will be closed.This issue will be monitored through the post market product monitoring review process.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 9618003.
 
Event or Problem Description
To date no additional patient or event details have been received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCORDION FLANGE
Common Device Name
PROTECTOR, OSTOMY
Manufacturer (Section D)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer (Section G)
CONVATEC DOMINICAN REPUBLIC INC
km 18.5 parque industrial
itabo, s.a. haina
san cristobal
Manufacturer Contact
pamela meadows
7815 national service road
suite 600
greensboro, NC 27409
3365424679
MDR Report Key9543865
Report Number9618003-2020-00095
Device Sequence Number16727961
Product Code EXE
UDI-Device Identifier00768455133710
UDI-Public00768455133710
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model Number421035
Device Lot Number8G04167
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2019
Supplement Date Received by Manufacturer09/01/2023
Initial Report FDA Received Date01/03/2020
Supplement Report FDA Received Date09/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device A
Patient Sequence Number1
Patient SexUnknown
-
-