C.R. BARD, INC. (COVINGTON) -1018233 BARD® SILASTIC® FOLEY CATHETER; SILASTIC FOLEY CATHETER Back to Search Results
Model Number 33616
Health Effect - Clinical Codes Pain (1994); No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
 
Event or Problem Description
It was reported that the silastic foley catheter was difficult to remove and was getting stuck in the urinary tract.The complainant reported that the catheter was cut in order for the balloon to be properly deflated, and the catheter was removed from the patient.The patient reportedly experienced pain.No medical intervention was reported.
 
Additional Manufacturer Narrative
The reported event was unconfirmed as the problem could not be reproduced.Nine unopened silastic catheters were returned.The sampling plan was used per general inspection.Therefore, three sample was chosen for evaluation.The 3 catheters were opened.The 3 catheters were inflated with 10ccs of air using a leur lock syringe.The catheters were passively deflated with no issues.The process was repeated with 10 ccs of water, there were no issues with inflation or deflation.A device history record review was not required per the investigation.The instructions for use were found adequate and state the following: "to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
It was reported that the silastic foley catheter was difficult to remove and was getting stuck in the urinary tract.The complainant reported that the catheter was cut in order for the balloon to be properly deflated, and the catheter was removed from the patient.The patient reportedly experienced pain.No medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD® SILASTIC® FOLEY CATHETER
Common Device Name
SILASTIC FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9544569
Report Number1018233-2020-00052
Device Sequence Number3444987
Product Code EZL
UDI-Device Identifier00801741020063
UDI-Public(01)00801741020063
Combination Product (Y/N)N
Initial Reporter StateNH
Initial Reporter CountryUS
PMA/510(K) Number
K951103
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 02/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number33616
Device Catalogue Number33616
Device Lot NumberNGCY3534
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/26/2019
Initial Date Received by Manufacturer 12/12/2019
Supplement Date Received by Manufacturer02/14/2020
Initial Report FDA Received Date01/03/2020
Supplement Report FDA Received Date02/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-