Model Number UCTW17
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Code Reaction (2414)
Date of Event 06/05/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiry date: 11/2021).
 
Event or Problem Description
It was reported that during an ultrasound-guided breast tissue marker placement procedure into hard tissue, the device allegedly was found to be potentially incompatible with the patient.It was further reported that upon test results two weeks post procedure, it was discovered the patient was allegedly allergic to a component of the device.A coaxial was used during the procedure and the procedure was completed by using the same marker.The patient was reportedly stable and will have the marker surgically removed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRACOR TWIRL MARKER
Common Device Name
BREAST TISSUE MARKER
Manufacturer (Section D)
BARD PERIPHERAL VASCULAR, INC.
1625 w 3rd st.
tempe AZ 85281
MDR Report Key9545998
Report Number2020394-2019-05898
Device Sequence Number11333548
Product Code NEU
UDI-Device Identifier00801741109416
UDI-Public(01)00801741109416
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K152510
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 01/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberUCTW17
Device Catalogue NumberUCTW17
Device Lot NumberHUCY2257
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2019
Initial Report FDA Received Date01/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age40 YR
-
-