Model Number 305901
Medical Device Problem Codes Device Damaged Prior to Receipt by User (1570); Material Twisted/Bent (2981)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a manufacturer representative regarding a trial patient who was using an external neurostimulator (ens).It was noted that the patient¿s trial started on (b)(6) 2019.It was reported that during the procedure, when attempting to place the second lead, it was determined that the tip of the lead was bent.It was noted that no environmental/external or patient factors were determined, as the lead appeared bent as soon as it was removed from its plastic sheath.They attempted to adjust the stylet within the lead without removing it completely, however, it did not change the significant bend in the lead.It appeared as if the lead tip was bent over on itself and that the stylet had pushed through the lead itself.A new lead was then opened and able to be successfully placed with no issues.There were no patient complications reported as a result of this event.
 
Additional Manufacturer Narrative
H6: analysis confirmed that the distal end of the lead was bent.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERSTIM
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE
Manufacturer (Section D)
AVAILMED S.A.DE C.V.9
c.industrial lt.001 mz.105
tijuana,bc 22444
MX  22444
MDR Report Key9546062
Report Number3010291427-2020-00001
Device Sequence Number2334735
Product Code EZW
UDI-Device Identifier00643169721166
UDI-Public00643169721166
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
P970004
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date07/29/2021
Device Model Number305901
Device Catalogue Number305901
Device Lot Number60204920
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/18/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2019
Supplement Date Received by Manufacturer03/31/2020
10/05/2020
Initial Report FDA Received Date01/03/2020
Supplement Report FDA Received Date04/08/2020
10/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age50 YR
-
-