| Model Number |
8888412007 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
12/23/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intraoperatively, on catheter insertion for capd (continuous ambulatory peritoneal dialysis), while checking the inflow of the device, they found a leak from the inner cuff site of the catheter.The catheter was not repaired, tego was not utilized, there was no luer adapter issue.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, intraoperatively, on catheter insertion for capd (continuous ambulatory peritoneal dialysis), while checking the inflow of the device, they found a leak from the inner cuff site of the catheter.It was stated that the lumens were flushed prior to use and nothing unusual was noted.Nothing else unusual was noted with the device prior to use and the packaging was stated to be normal.The catheter was not repaired, tego was not utilized, there was no luer adapter issue.A new catheter was inserted and used to resolve the issue and complete the procedure.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d10, g4, h3, h6 h3 evaluation summary: post market vigilance (pmv) led an evaluation of one device.The visual inspection of the returned product noted: the cannula, cuffs, adapter, clamp and plug appeared intact.The adapter was inserted into the cannula and the catheter was submerged into a water bath.The end was clamped, and a syringe was used to inject air to observe leakage.Air bubbles were present at the distal end of the cannula near the inner cuff.A hole was observed.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the hole in the cannula may occur when contact is made with a sharp surgical instrument during clinical application.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|