Model Number K287
Device Problem Failure to Capture (1081)
Patient Problems Complaint, Ill-Defined (2331); No Code Available (3191)
Event Date 10/13/2019
Event Type  Injury  
Manufacturer Narrative
As no further investigation is complete.This investigation will be updated should further information be provided.
 
Event Description
It was reported that the patient implanted with this pacemaker reported feeling unwell and complained of hot flashes.Interrogation revealed evidence of intermittent loss of capture (loc).The right ventricular (rv) lead output was reprogrammed and the patient will be monitored.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VITALIO
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key9546572
MDR Text Key173565306
Report Number2124215-2019-23813
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/26/2017
Device Model NumberK287
Device Catalogue NumberK287
Device Lot Number247528
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/13/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/01/2015
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Required Intervention;
Patient Age89 YR
-
-