RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problem Increase in Pressure (1491)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2019.Date of report: 01/03/2020.
 
Event Description
The customer reported high o2 (oxygen) pressure, and no o2 available.There was no patient involvement.
 
Manufacturer Narrative
G4: 15jan2020.B4: (b)(6).The manufacturer¿s field service engineer (fse) could not replicate the problem.Replaced all parts ordered.Performed full testing.All tests passed.No fault found.(est) extended self test passed.The device was not being used for treatment when the reported event occurred, and there is a relationship of the device to the reported problem.No parts returned to failure investigation; therefore, the root cause of the reported issue could not be determined.If parts are returned, the complaint will be reopened and an investigation will be performed.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
G4: 25feb2020 b4: (b)(6)2020.The oxygen (o2) solenoid and daq board were returned to the manufacturer for failure investigation (fi).After testing, it was determined that the customer complaint was not verified.There was no fault found.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key9546856
MDR Text Key181426071
Report Number2031642-2020-00022
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 12/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2020
Was the Report Sent to FDA? No
Date Manufacturer Received12/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-