• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION PENTA 3MM LEAD, 60 CM; SCS PADDLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION PENTA 3MM LEAD, 60 CM; SCS PADDLE LEAD Back to Search Results
Model Number 3228
Device Problems Disconnection (1171); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 11/11/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Related manufacturer report number: 3006705815-2020-00017.It was reported that the patient recently had a fall, after which they began experiencing auto reducing and the lead was found to have high impedances.X-ray showed a possible lead pullout.In turn, surgical intervention may take place to address the issue.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
Additional information revealed that the patient underwent surgical intervention wherein the lead was explanted and replaced.Postoperatively, the patient has effective therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENTA 3MM LEAD, 60 CM
Type of Device
SCS PADDLE LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9547363
MDR Text Key173585457
Report Number1627487-2020-00014
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05414734406246
UDI-Public05414734406246
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date07/28/2016
Device Model Number3228
Device Catalogue Number3228
Device Lot Number4687160
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/18/2019
Initial Date FDA Received01/03/2020
Supplement Dates Manufacturer Received03/18/2020
Supplement Dates FDA Received04/02/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/08/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
3789, SCS IPG
Patient Outcome(s) Other;
Patient Age40 YR
Patient Weight91
-
-